Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of three Standard reps.
Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.
Cornerstone氫氧氮分析軟件同時(shí)支持***的書記文檔格式和可變的報(bào)告格式。江蘇LECO分析儀哪家好
美國力可LECO氧氮?dú)浞治鰞x
公司介紹:
美國力可公司(LECO)始于1936年 ,以生產(chǎn)用于冶金、機(jī)械、能源、化工、農(nóng)業(yè)、食品及科研開發(fā)領(lǐng)域的各種分析儀器。
ONH836氧氣/氮?dú)?氫氣元素分析儀設(shè)計(jì)用于使用惰性氣體熔融技術(shù)對(duì)無機(jī)材料,鐵和非鐵合金以及耐火材料中的氧,氮和氫含量進(jìn)行寬范圍的測(cè)量。
堅(jiān)固的樣品加載區(qū)域可簡化日常維護(hù)并大程度減少大氣污染
用于氧氣檢測(cè)的三重紅外電池可實(shí)現(xiàn)業(yè)內(nèi)大的檢測(cè)范圍
通過自動(dòng)化選項(xiàng)提高實(shí)驗(yàn)室生產(chǎn)率
可用的自動(dòng)清潔器大程度地減少了兩次分析之間的手動(dòng)清潔需求
適用于坩堝和樣品的20位往復(fù)式裝載機(jī)
安裝在動(dòng)臂上的觸摸屏界面可提供更好的人體工程學(xué)設(shè)計(jì)和直觀的操作
紅外(IR)和熱導(dǎo)率(TC)檢測(cè)器,沒有活動(dòng)部件,也無需手動(dòng)調(diào)節(jié)
改善TC電池的穩(wěn)定性
可繞過的OMI-2載氣洗滌塔,提高了TC電池的穩(wěn)定性
獲得Dyanamic流量補(bǔ)償可進(jìn)一步增強(qiáng)TC細(xì)胞的穩(wěn)定性
江蘇LECO分析儀哪家好可選具有壓力補(bǔ)償和溫度控制性能的標(biāo)氣檢測(cè)氣路。
美國LECO力可氧氮?dú)浞治鰞xONH836 氧氮?dú)渎?lián)測(cè)儀系列軟件部分
1、用戶友好界面的Cornerstone軟件
力可**的Cornerstone氧氮?dú)浞治鲕浖捎糜|摸屏界面使用戶可完整的達(dá)到分析控制、方法設(shè)定、在線診斷、數(shù)據(jù)報(bào)告。Cornerstone氧氮?dú)浞治鲕浖试S用戶在單一分析界面中完成日常的所有工作,快速而且簡單易用。力可創(chuàng)新的批次和重復(fù)樣品數(shù)據(jù)分組功能極大簡化了樣品結(jié)果輸出和結(jié)果標(biāo)準(zhǔn)偏差的自動(dòng)計(jì)算。減少了繁瑣的額外數(shù)據(jù)處理。
2、軟件特點(diǎn)和優(yōu)勢(shì)
Cornerstone氧氮?dú)浞治鲕浖殖伤膫€(gè)主要部分:分析、診斷、設(shè)定、儀器,界面可快速導(dǎo)航和組織。工具條、滑動(dòng)條和下拉式菜單簡化了校正和數(shù)據(jù)處理的參數(shù)設(shè)定。分析軟件同時(shí)包括實(shí)時(shí)的環(huán)境參數(shù)的檢測(cè),提供完整的系統(tǒng)圖表模擬。高級(jí)診斷程序包括全數(shù)字化的在線操作手冊(cè)、維護(hù)模擬演示、圖片說明等提供快速指導(dǎo)。Cornerstone氧氮?dú)浞治鲕浖瑫r(shí)支持***的數(shù)據(jù)文檔格式和可變的報(bào)告格式。
Procedure – Powder/Chip Samples
Prepare the instrument as outlined in the operator's instruction manual.
Login a minimum of three Blank reps.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
采用三個(gè)**的紅外器檢測(cè)氧含量,同時(shí)檢測(cè)CO和CO2,然后在轉(zhuǎn)化為CO2集中檢測(cè)。上海元素分析儀咨詢報(bào)價(jià)
提供各種無機(jī)材料、鋼鐵、有色金屬、硬質(zhì)合金、陶瓷材料中寬范圍的氧氮?dú)渫瑫r(shí)分析。江蘇LECO分析儀哪家好
氧氮?dú)浞治鰞x的常見故障及解決方法:
氧和氮空白值超過20。這是由于氣流小,不能將爐子中的空氣驅(qū)趕出去??烧{(diào)節(jié)氣體流量,調(diào)節(jié)載氣壓力在0.2~0.4MPa。接通儀器載氣,放一個(gè)石墨坩堝在下電極上,打開主電源開關(guān),點(diǎn)擊軟件上的關(guān)爐按鈕,關(guān)閉爐子并等待10s。調(diào)節(jié)流量調(diào)節(jié)器,直到流量計(jì)a顯示為30L/h,打開爐子。調(diào)節(jié)調(diào)節(jié)器直到流量計(jì)b顯示為50L/h,再次關(guān)閉爐子。如果以上設(shè)置不穩(wěn)定,則增加流量至100L/h,反復(fù)調(diào)節(jié)直至儀器穩(wěn)定。 江蘇LECO分析儀哪家好